Supplied with one BT1603-B 12 x 12mm hole block. Additional blocks are available (See Related Products).
A gas flow control valve is also available for precise control of gas flow.
Features
- Works by heating with dry bath in the bottom and nitrogen blowing on the surface which accelerates liquid evaporation and sample concentration
- Adjustable position of gas needle in air chamber plate for different types of tubes. The length of a standard gas needle is 150mm
- Adjustable height of air chamber plate. Sample surface in the tube can be observed while concentrating, a unique characteristic
- Independently controlled gas needle. Separate blowing and flow regulating of each needle are available to avoid gas waste
- The instrument can be put into a ventilation cabinet when the concentration sample is toxic solvents
- Built-in overheating protection, automatic fault detection and fault alarm features
- LED displays temperature and time. Operation is simple and convenient. Can be set for continuous operation.
- Works by heating with dry bath in the bottom and nitrogen blowing on the surface which accelerates liquid evaporation and sample concentration
- Standard configured air cavity and adjustable bracket.
- Optional blocks are available.
Specifications
- Temp control range: ambient+5°C-150°C
- Temp setting range: 5°C-150°C
- Temp stability @ 40-100°C:≤±0.5 °C
- Temp stability @ 100-150°C: ≤±1 °C
- Block temp uniformity@100°C:≤±0.5 °C
- Block temp uniformity@150°C:≤±1 °C
- Temp display accuracy: 0.1°C
- Heating speed: ≤30 min (40°C to 150°C)
- Time range: 1min-99h59min or continuous
- Needle plate max lift stroke: 285mm
- Gas-in joint outer diameter: ɸ7mm
- Nitrogen pressure: ≤0.1MPa
- Nitrogen flow rate: 0-10L/min
- Needle length: 150mm
- Sample capacity: 1 standard visible block
- Voltage: AC110V, 50/60Hz
- Power: 400W
- Fuse: 250V,3A/6A, ɸ5x20
- Dimension (WxDxH): 220x260x 525mm
- Weight: 5.8kgs